Vitalograph 3 Litre Calibration Syringe

Vitalograph precision syringes provide quick and easy accuracy checks with reliability and fast verification.

Providing assured accuracy for calibration verification and linearity checking.

627.63 ex VAT

SKU: A101553Categories: ,

Shipping 1-2 Business Days

Trading for over 25 Years

The 3L precision syringe can be used with flow or volume sensor systems and is compatible with any manufacturers’ equipment.  Each syringe has a unique serial number and history file with accuracy certified to better than 0.5%.

For flow measuring spirometers, the 2019 ATS/ERS Standardization of Spirometry guidelines recommends daily calibration verification using a 3 liter syringe with a filter attached.

The 3-L precision syringe is recommended for all spirometers as required by ISO 26782:2009.  A range of adaptors is available.

Compatible with:

  • Carefusion (Micro Medical) spirometers
  • Vitalograph spirometers
  • MIR spirometers
  • NDD Easy On-PC spirometer

Specifications:

  • Certified accuracy better than 0.5%.
  • Compatible with all spirometry equipment; a range of adaptors is available.
  • Simple end-to-end volume, no lining up marks or grooves.
  • Unique serial number and history file for each syringe.
3L Precision Syringe
Product 3L Precision Syringe Cat No 36020
Model 2040
Gross Size 490mm x 160mm x 125 mm
Gross Weight 2.5kg

The 2019 ATS/ERS Standardisation of Spirometry guidelines recommend:

“Calibration verifications must be undertaken at least daily using a 3-L syringe cycled at least three times to give a range of flows varying between 0.5 and 12 L/s (with 3-L injection times between 0.5 and 6 s). If an in-line filter is used in spirometry testing, then it must also be used during recalibrations and verifications.”

Yes. The 2019 ATS/ERS Standardisation of Spirometry guidelinesrecommend:

“A biological control is not a substitute for the use of a calibration syringe. However, operators are encouraged to know their own usual FEV1 and FVC, which allows them to conduct a quick, rough check if they suspect a problem.”